9-3
Troubleshooting
Operator Manual
P129394-176
9.2.2 Too Long In
Exhaust
Occurs if chamber
does not exhaust to
4 psig within the
allotted time.
Screen with Service Instructions
Screen with Operator Instructions
9.2.3 Too Long In
Evacuation
Occurs if chamber
does not reach the set
evacuation level within
the allotted time.
STATUS . . ALARM!
200
TOO LONG IN EXHAUST
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPETE CYCLE
•
EXTEND EXHAUST TIME
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
201
TOO LONG IN EXHAUST
-> CHAMBER DID NOT EXHAUST TO ATMOSPHERIC
PRESSURE WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. CHAMBER DRAIN STRAINER PLUGGED
-> CLEAN
2. SOLENOID VALVE MALFUNCTION
-> REPAIR S03
3. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
PAPER
FEED
SERVICE
HELP
STATUS . . ALARM!
202
TOO LONG IN EVACUATION
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPETE CYCLE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CHECK WATER SUPPLY VALVE
-> IF CLOSED, OPEN VALVE
3. IF ALARM RECURS, ABORT CYCLE
AND CALL SERVICE
SERVICE
HELP
ABORT
SILENCE
ALARM