1-2
P129394-176
Operator Manual
Safety Precautions
WARNING – EXPLOSION HAZARD:
WARNING – SLIPPING HAZARD:
WARNING – PERSONAL INJURY HAZARD AND/OR EQUIPMENT DAMAGE HAZARD:
Allow sterilizer to cool to room temperature
before performing any cleaning or maintenance
procedures.
Failure to shut off the steam supply
when cleaning or replacing strainers can result in serious injury.
Jacket pressure must be 0 psig before beginning work on the steam trap.
Proper testing of the safety valve requires the valve to be operated under pressure.
Exhaust from
the safety valve is hot and can cause burns. Proper safety attire (gloves, eye protection, insulated overall)
as designated by OSHA, is required. Testing is to be performed by qualified service personnel only.
Steam may be released from the chamber when door is opened.
Step back from the sterilizer each
time the door is opened to minimize contact with steam vapor.
This sterilizer
is not designed to process flammable compounds.
To prevent falls, keep floors dry
by immediately wiping up any spilled liquids or condensation in
sterilizer loading or unloading area.
Regularly scheduled preventive maintenance is required for safe and reliable operation of this equipment.
Contact your STERIS service representative to schedule preventive maintenance.
When closing the chamber door, keep hands and arms out of the door opening and make sure opening is
clear of obstructions.
Repairs and adjustments to this equipment must be made only by fully qualified service personnel.
Maintenance performed by inexperienced, unqualified persons or installation of unauthorized parts could
cause personal injury, result in costly equipment damage and/or void the warranty.
The performance of the 510k cleared AMSCO C Series Sterilizer is validated as a system including
components defined by STERIS in the Operator Manual and Service Manual for the AMSCO C Series
Sterilizer. Substitution of unauthorized components can potentially lead to personal injury, damage or
premature failure of the product and result in a unit configuration that is inconsistent with the validated,
FDA cleared product. Altering the equipment or using unapproved component’s may void STERIS’
warranty. DO NOT USE components that are not validated as part of the AMSCO C Series Sterilizer.
WARNING – STERILITY ASSURANCE HAZARD:
Load sterility may be compromised if the biological indicator or air leak test indicates a potential problem.
If these indicators show a potential problem, refer the situation to a qualified service technician before
using the sterilizer further.