Routine Maintenance
2.
Wash the inside of the chamber and shelf assembly (plus any
other loading equipment) with a mild detergent solution such as
Liquid-Jet 2 Instrument Detergent or current equivalent product.
(
Contact STERIS.)
3.
Once the chamber is clean, replace the shelf assembly using the
reverse of the appropriate procedure given in
Step 1
. Close the
chamber emergency manual exhaust valve.
4.
Professional cleaning of the chamber on a yearly basis (or as
local conditions require) is suggested to maintain appearance of
the chamber interior. Contact STERIS for information regarding
this service.
CAUTION – POSSIBLE
EQUIPMENT DAMAGE:
•
Lifting the chamber float
switch when cleaning the
chamber may cause the
sterilizer control to initiate a
“Chamber Flooded” alarm. If
this alarm condition occurs,
the operator must turn the
control power OFF then ON
to clear the alarm. The
control power switch is
located in the mechanical
area at the side of the
sterilizer. Placing the
sterilizer in standby does not
clear this alarm.
•
Never use a wire brush,
abrasives, or steel wool on
door and chamber assembly.
Do not use cleaners containing
chloride on stainless-steel
surfaces. Chloride-based
cleaners will deteriorate
stainless steel, eventually
leading to failure of the vessel.
!