background image

F-1.

General

This section of the manual contains routine 
maintenance information regarding the AMPTEC 
RESEARCH 630ES Igniter tester.  Calibration 
should be performed on a regular basis to ensure 
continued instrument accuracy or following a main 
PCB electronic component repair / replacement. The 
recommended calibration interval is 1 year. 

The AMPTEC 630ES  Igniter Tester  is a four wire 
Kelvin ohmmeter. 

The AMPTEC 

 Igniter 

Tester  must be calibrated using four wire Kelvin 
connections to the resistance standard

 in order to 

eliminate lead resistance and contact resistance 
errors. The 

Option “630-304” 

is a Gold Plated 4 

Wire Kelvin ( or equivalent) Banana Jack Test Lead 
Set that should be used  to calibrate the AMPTEC  
630ES Igniter Tester. The 4-wire configuration is 
maintained up to the point of the connection to the 
resistance standard. 

F-2.

Required Test Equipment

Following standard resistors are required to calibrate 
the 630ES  Igniter Tester  .    

Precision Resistors:

0.001 ohm ± 0.01% Accuracy (2 Ohm 

zero

)

0.1 ohm ± 0.01% Accuracy  (20-20K Ohm 

zero

)

1.0 ohms ± 0.005% Accuracy (2 ohm

 fullscale

)

10 ohms ± 0.005% Accuracy (20 ohm 

verify

)

100 ohms ± 0.005% Accuracy (200 ohm 

fullscale

)

100 KOhm ± 0.005% Accuracy (200 Kohm 

verify

)

Test Leads:

4-wire Kelvin Test Lead set or AMPTEC

 

Option "630-304" Kelvin 4 Wire Banana Test Leads

F-3.

Calibration Procedure

The AMPTEC

 

630ES should be calibrated with fully 

charged batteries 

(NICAD) 

and should be allowed to 

warm-up

 for a minimum of 15

  

630ES

minutes

 before

beginning the calibration procedure.  Some 
calibration adjustments are accessed by removing 
the calibration access screws (

cont next column

)

in the top-plate perimeter of the unit. Use the 
two handles (careful not to break any umbilical 
cabling to the main PCB) to lift up the top plate 
to access most of the main PCB RV trimpots.  
The locations of the adjustments are shown on 
drawing number 630-600 at the back of this 
manual.

F-3-1. Zero Offset Adjustment

 (20 ohm and 

higher ranges)

1. 

Select the 20 ohm range.  Connect the Kelvin 

leads to the 0.1 ohm standard resistor.

2. 

Adjust potentiometer RV2 (

) for a display indication of 00.10 

Ohms . 

Do not over adjust RV2 past a  0.00 

reading

. A false or negative polarity 0.010 

display reading offset error can  be created. This 
calibration error has the display appearing 
normal (even though a negative 0.010 offset has 
been created ) - the display doesn't indicate a 
negative sign).

F-3-2. Full Scale Adjustment 

 (20 ohm and 

higher ranges)

1. 

Select the 200 ohm range.  Connect the 

Kelvin leads to the 100 ohm standard resistor.

2. 

Adjust RV1 for a display reading of 100.00.

1. 

Select the 2 ohm range.  Connect the 4 wire 

Kelvin leads to the 0.001 ohm standard resistor.

2. 

Adjust potentiometer RV5 (on the main PCB) 

for a display indication of 0.0010 Ohms . 

Do 

not over adjust RV5 past a  0.00 reading

. A 

false or negative polarity 0.0010 display reading 
offset error can  be created. This calibration 
error (

mis-adjusted beyond zero

) has the display 

appearing normal (even though a negative 
0.0010 offset has been created ) - the display 
doesn't indicate a negative sign)

the front panel 

zero pot located on the top plate - not found on 
the main PCB

F-3-3     2.0 Ohm Range Zero adjustment  

Summary of Contents for 630ES

Page 1: ...RESEARCH AMPTEC PHONE 512 858 4045 800 350 5105 website http www amptec com FAX 512 858 4340 TOLL FREE IN USA Revision D AMPTEC 630ES SAFETY IGNITER TESTER OPERATION MAINTENANCE MANUAL...

Page 2: ...ry Charger Power Requirements A 5 Setup and Use SECTION B 630ES Explosive Safety Igniter Tester SPECIFICATIONS Table B 2 Specifications SECTION C REPLACEMENT OPTIONAL AND ACCESSORY ITEMS C 1 Available...

Page 3: ...y Scheme Figure E 1 Model 630ES Block Diagram Figure E 1B Analog to Digital Converter Timing Diagram Figure E 2 Analog Section of IC1 and IC2 Figure E 3 Constant Diagram Simplified Figure E 4 LED Disp...

Page 4: ...it useful in a variety of applications Please check the front of this manual for any addendums that may apply to new 630ESs and 630AN conversions SECTION A UNPACKING AND INSPECTION A standard feature...

Page 5: ...630ES can be opened If the Purge Valve was opened return the valve to the closed state tighten clockwise once the 630ES lid is open The 630ES has a way to remove it s lid completely for example when u...

Page 6: ...ng 220VAC 50 HZ AC line power for running the battery charger please inform AMPTEC s sales department at the time of your order and the appropriate adaptor will be included with the 630ES Failsafe Ohm...

Page 7: ...ge clamped at 1 6 volts 2 Power 4 D 5 0 AHr Heavy Duty 1 2V rechargeable nickel cadmium batteries Fuse 2 ampere fast blo 3 AG type AC DC Battery Charger Option 630DC provides 9VDC at 1 1 Ampere nomina...

Page 8: ...d set is terminated with the mating connector for 630ES Front panel connection port user specified AMPTEC RESEARCH s website http www amptec com should also provide the latest information on accessori...

Page 9: ...nd phone changes information C 1 Available Accessories and Options This manual does not list all possible accessories that AMPTEC RESEARCH is willing to provide as a support items for the 630BN Ignite...

Page 10: ...tip Option 630 304 Gold 4 Wire Kelvin Clip Test Lead Set Option KCS Kelvin Clips accept banana jacks in back works with Option 630 403 is a 4 wire Kelvin Handheld Probe Lead set terminated with two ha...

Page 11: ...or the 20 Ohm lowest resistance range is just below the display on the left see item 20 of the 630ES Front Panel Diagram When a given range is selected an indicator LED informs the user Also note that...

Page 12: ...ion AMPTEC 630ES Front Panel Diagram 1 Calibration Maintenance access screws also can be easily protected with a tamper proof calibration sticker or seal 2 Purge Valve provides air pressure equalizati...

Page 13: ...13 10 ohm test resistor for checking the mid scale functionality of the 20 ohm range 14 Input jack for connecting to the 630ES Ohmmeter common test lead plug in black banana jack This common jack is a...

Page 14: ...stepped down from the 9VDC 1 1A supplied by the charger The correct charging voltage is supplied by the adapter included with the instrument Additional AC DC Battery Chargers are available as Option 6...

Page 15: ...keyed test lead connector is optionally available for customized wiring connections kelvin clips cables terminated with spade lugs and special banana jacks All AMPTEC ohmmeters use a high impedance v...

Page 16: ...ng the test resistors in the Functional Test Section then the connect ion problem must be outside of the 630 series meter i e your wiring harness or the actual device under test connection If the 630...

Page 17: ...the fuse from the fuseholder With an adjustable DC power supply set the power supply output to be 4 50 VDC Be sure to observe power supply polarity Connect the power supply to the test points labeled...

Page 18: ...te the batteries need charging or fuse needs replacing If the 630ES exhibits problems that cannot be eliminated by reviewing Chapters B and D the following guidelines have been established to help sol...

Page 19: ...or move the replacement part until the solder has cooled Cold solder and bad solder joints can cause more problems Use the chassis ground connect to the common terminal of the functional test section...

Page 20: ...03 takes care of the digital part of the 4 digit 20 000 count dual slope conversion COUNTERS DISPLAY COMPARATOR BUFFER MULTIPLEXER ZERO CROSSING DETECTOR INTEG UUT CONTROL LOGIC INPUT SWITCH DC CONSTA...

Page 21: ...EGRATOR INTEGRATOR COMPARATOR COMPARATOR CAZ CAZ CSTRAY CSTRAY RINT RINT FIGURE E 2B PHASE ll INTEGRATE INPUT FIGURE E 2C PHASE lll AND DE INTEGRATE CINT CINT VIN VIN CREF CREF 3 3 1 1 2 2 6 6 5 5 4 4...

Page 22: ...ished in Phase I The time or number of counts required to do this is proportional to the input voltage E 4 1 Reference Voltage The precision reference voltage required to do the A D conversion is deve...

Page 23: ...connect the 5 volt supply directly across R218 D202 the range resistor and R L D203 however acts as a 1 6 volt zener diode limiting the voltage that can appear across these components Even if every co...

Page 24: ...Power Supply Scheme The 5 volt power supply is provided directly by the batteries for driving the LED displays and digital logic The 5VD is used for driving IC8 the low battery detection circuit The 1...

Page 25: ...um of E plus the output of in IC202 Consider now that the short is removed from the Ihi and I terminals and a 100 ohm resistor R is lo L connected in its place The current through RL increases the vol...

Page 26: ...hm range RL2 is for the 200 Ohm range RL3 is for the 2K Ohm range and RL4 is for the 200K Ohm range A B G F C E D D P SEGMENT PIN A 1 B 13 C 10 D 8 E 7 F 2 G 11 D P 6 Common 14 PIN 1 2 3 4 5 6 7 PIN 1...

Page 27: ...g the calibration procedure Some calibration adjustments are accessed by removing the calibration access screws cont next column in the top plate perimeter of the unit Use the two handles careful not...

Page 28: ...2 Ohm Range Full Scale Adjustment 1 Select the 2 ohm range Connect the 4 Wire Kelvin leads to the 1 00 ohm standard resistor 2 Adjust RV6 on main PCB for a display reading of 1 0000 Ohms 3 Check the 2...

Reviews: