2. Once 30 seconds has passed, the system will advance automatically to the next screen. Follow
the on-screen instructions and apply the Calibration Cup to the detector. The measurement will
begin when the Calibration Cup is detected and the screen will advance automatically.
Figure 31: Response Test - Apply Cup
NOTE
As the gas concentration increases, setpoint flags may become set,
triggering a status message.
3. The timer screen will be shown, and the gas concentration measurement will start to increase.
Once 90% of the target value has been reached the screen will advance.
Figure 32: Response Test - Timer
4. Once 90% of the target value has been reached, the screen will be displayed. In this example the
response time was 24 seconds. Once the result has been recorded, remove the Calibration Cup
and press the centre of the keypad to leave.
Figure 33: Response Test - Result