128797 05
2017-08-11
65
H120(90)672
0,1
0,01
-80
260
-112
500
0,1
0,18
P100(90)JIS
0,1
0,01
-200
100
-328
212
0,14
0,26
100
500
212
932
0,22
0,4
YSI(90)400
0,1
0,01
15
150
59
302
0,02
0,04
Read accuracy is based on 4 wire input.
Source accuracy in terminals 2 wire source.
Pressure modules, Barometric option (BARO) and APM CPF
APM CPF Type
(s)
Gauge
12 month
Accuracy ±
0 to 30 % range
12 month
Accuracy ±
30 to 110% range
12 month
Accuracy ±
Vacuum % FS
Bar
MPa
psi
3 bar
300 kPa
30 psi
-1
3
-0.099 0.300
-14.5
30
0.0075% FS
0.025% RDG
0.06% FS
10 bar
1 MPa
100 psi
-1
10
-0.099
1.0
-14.5
100
0.0075% FS
0.025% RDG
0.06% FS
30 bar
3 MPa
300 psi
-1
30
-0.099
3.0
-14.5
300
0.0075% FS
0.025% RDG
0.06% FS
100 bar
10 MPa
300 psi
0
100
0
10.0
0
1 000
0.015% FS
0.05% RDG
N/A
300bar
30 MPa
3 kpsi
0
300
0
30.0
0
3 000
0.015% FS
0.05% RDG
N/A
700 bar
70 MPa
10 kpsi
0
700
0
70.0
0
10 000
0.03% FS
0.1% RDG
N/A
1000 bar
100 MPa
15 kpsi
0
1000
0
100.0
0
15 000
0.03% FS
0.1% RDG
N/A
Absolute pressure
APM CPF with ASC-400 BARO option / 12 month Accuracy ±
3 bar APM CPF
Accuracy
±
300 kpA APM CPF
Accuracy
±
30 psi APM CPF
Accuracy
±
0.0138 to 1 barA
0.0008 barA
1.38 to 100 kPaA
0.08 kPaA
0.2 to 14.5 psiA
0.011 psiA
1 to 4 barA
0.025% RDG +
0.0003 barA
100 to 400 kPaA
0.025%
RDG +
14.5 to 44.5 psiA
0.025% RDG
+ 0.003 psiA
10 bar APM CPF
Accuracy
±
1 MPa APM CPF
Accuracy
±
100 psi APM CPF
Accuracy
±
0.0138 to 1 barA
0.0008 barA
0.00138 to 0.1 MPaA
0.00008 MPaA 0.2 to 14.5 psiA
0.011 psiA
1 to 4 barA
0.001 barA
0.1 to 0.4 MPaA
0.0001 MPaA
14.5 to 44.5 psiA
0.011 psiA
4 barA to 11 barA
0.025% RDG
0.4 MPaA to 1.1 MPaA 0.025% RDG
44.5 to 114.5 psiA 0.025% RDG
30 bar APM CPF
Accuracy
±
3 MPa APM CPF
Accuracy
±
100 psi APM CPF
Accuracy
±
0.014 to 1 barA
0.001 barA
0.0014 to 0.1 MPaA
0.001 MPaA
0.2 to 14.5 psiA
0.01 psiA
1 to 10 barA
0.003 barA
0.1 to 1.0 MPaA
0.003 MPaA
14.5 to 104.5 psiA 0.03 psiA
10 barA to 31 barA 0.025% RDG
1.0 MPaA to 3.1 MPaA 0.025% RDG
104.5 to 314.5 psiA 0.025% RDG
Specified temperature range -10 to 50°C / 14 to 122°F (APM CPF & BARO option) Vacuum FS, 1 bar / 100 kPA
/ 14.5 psi. F.S. (full scale) is the numerical value of the positive pressure range. Accuracy includes hysteresis,
nonlinearity, repeatability and reference standard uncertainty, 1 Year typical longterm stability, operated
inside the rated temperature span and pressure range. Requiring frequently zeroing.