128797 05
2017-08-11
46
3. Connect the pump to the ASC-400 calibrator and the pressure switch.
4. Make sure the vent on the pump is open. Zero the pump if necessary.
Note…
Zero the APM if necessary. Close the vent after zeroing the calibrator.
5. The upper display shows no read outs at neither “Closed”, “Opened” nor “Dead
band”
6. Apply pressure with the pump slowly until the switch opens.
Note…
In the switch test mode the display update rate is increased to help capture
changing pressure inputs. Even with this enhanced sample rate pressurizing
device under test should be done slowly to ensure accurate readings.
7. Vent the pump slowly until the pressure switch closes.
8. The upper display will now show :
the pressure when the switch closed
the pressure that the switch opened at
the dead band between the pressure readings.