128797 05
2017-08-11
32
6.6
Sourcing mV
To source an mV, follow the principle of navigating through the functions described in
the guidelines in section 4.2 and proceed as follows
:
1. Connect the thermocouple leads to the appropriate polarized TC mini plug, and
insert the plug into the TC terminals on the calibrator, as shown in Figure 10.
2. Select “Voltage” from the Lower Mode list.
3. Select “-10 to 75 mV” from the Range list.
4. Select “Out” from the Test Mode list.
5. Enter the desired mV using either the ARROW keys or the numeric keys.
Note…
CJ compensation is not active in this mode.