128797 05
2017-08-11
37
3. If TC or RTD is selected, choose the type and temperature units first, to get proper
step values.
4. Selection of Step is also possible using the function key (Step) for quick,
menu free access.
5. Select the desired step size from the Step Size list. The ASC-400 will step from the
0% set source value to the 100% set source value in 10%, 20% or 25%increments.
The step time is adjustable from 1 to 999 seconds.
6. Exit the edit mode menu when all the selections have been made.
7. Press (Step) to access the MANUAL Step menu.
or
8. Press (Auto) to start auto stepping, using the selected step time.
9. Pres either (–Step) or (+Step) to Step up- or downwards.
10. In AUTO step, press (Stop) to stop the auto stepping.
11. To cancel Step or Auto step mode, press (Back).
6.10.2 Using the Ramp function
To use the Ramp function, follow the principle of navigating through the functions
described in the guidelines in section 4.2 and proceed as follows:
1. Select the desired lower mode from the Lower Mode list. Auto Output options are
not available for Pulse, Pressure and Baro modes.
2. Select “Out Ramp” from the Test Mode list.
F3
F2
F3
F3
F4
F4
F1