
Controller/Interface
|
2-19
PN 9000-165-VE, Rev P
7. When you have entered all of the values, press
ENTER/MENU
to
START
the calibration.
8.
Inject the Oxygen Span Gas
You are prompted to inject the O
2
span gas into the calibration gas inlet
port on the sensor (tubing must be free of oil and dirt). Press
ENTER/
MENU
when you have applied the span gas.
9. The screen displays the status of the O
2
span gas as “
Wait
” or “
Ready
.”
When the status is “
Ready
,” press
ENTER/MENU
to accept the values.
If you accept the values before the status is “
Ready
” you will be asked
to acknowledge that the reading is not stable before continuing.
10.
Inject Zero Gas
You are then prompted to apply the O
2
zero gas to the calibration inlet port of
the sensor. Press
ENTER/MENU
when you have applied this gas.
11. The system will calibrate the O
2
and combustibles zero gas. When the read-
ings have stabilized (Ready), press
ENTER/MENU
to accept the readings.
Figure 2-13.
Calibration screen gas status
reading.
Summary of Contents for AMEVision
Page 78: ...PN 9000 165 VE Rev P WDG V AMEVision Modbus Map A 2 Thermox AMEVision Display User Interface ...
Page 79: ...PN 9000 165 VE Rev P EventMask 1 Bit definitions AMEVision Modbus Information A 3 ...
Page 80: ...PN 9000 165 VE Rev P WDG 5 Command Definitions A 4 Thermox AMEVision Display User Interface ...
Page 81: ...PN 9000 165 VE Rev P Status Mask Bit definitions AMEVision Modbus Information A 5 ...
Page 83: ...PN 9000 165 VE Rev P EventMask 2 Bit definitions AMEVision Modbus Information A 7 ...