PN 9000-165-VE, Rev P
AMEVision Host Display Mains Supply Connections
The AMEVision Host Display can operate using between 85–250 volts AC,
47–63 Hz. There is no power switch or circuit breaker on the control unit, and
it must be protected by installing it on a circuit-protected line, maximum 15
amperes, with a switch or circuit breaker in close proximity to the control unit
and within easy reach of an operator. Mark the switch or circuit breaker as the
control unit disconnecting device.
Mains supply connections to the control unit are as follows:
• Line – Line connection
• Neutral – Neutral connection (USA)
• Chassis Stud – Equipment ground (protective conductor)
Use the 1/2" conduit entry hole for AC mains supply wiring. Use the chassis
stud next to the 1/2" conduit entry hole for equipment ground (protective
conductor).
AC
(
L
) and (
N
) markings are provided by the terminal block for connection of
AC power. These markings are for reference purposes only, such as for use on
system wiring diagrams, etc. The system/product has or needs no specific
LINE
or
NEUTRAL
connection for any function, safety or otherwise. The (
N
) terminal
is not internally grounded, nor needs to be. The system will operate normally
regardless of what AC input terminal (
L
or
N
) the
AC Line
or
Neutral
is con-
nected to, or, if there is a
Neutral
used at all (i.e., 208 VAC US power connec-
tion).
Overview and Installation
|
1-3
Summary of Contents for AMEVision
Page 78: ...PN 9000 165 VE Rev P WDG V AMEVision Modbus Map A 2 Thermox AMEVision Display User Interface ...
Page 79: ...PN 9000 165 VE Rev P EventMask 1 Bit definitions AMEVision Modbus Information A 3 ...
Page 80: ...PN 9000 165 VE Rev P WDG 5 Command Definitions A 4 Thermox AMEVision Display User Interface ...
Page 81: ...PN 9000 165 VE Rev P Status Mask Bit definitions AMEVision Modbus Information A 5 ...
Page 83: ...PN 9000 165 VE Rev P EventMask 2 Bit definitions AMEVision Modbus Information A 7 ...