Maintenance
SPECTRO ARCOS FHX3X — 08.02.2022
— 47
8. Put the nebulizer into a beaker and carefully fill it with diluted hydrochloric
acid.
Hazardous substance!
Observe the corresponding information on hazardous materials.
9. Let the acid act for up to 15 minutes (depending on the degree of
contamination).
When heavy contamination must be removed, another suitable acid or
aqua regia may be required.
If there are deposits on the capillary tube, remove them very carefully
with a fine copper wire. Extreme care is required in order to avoid
damage to the capillary tube.
10. Then flush the nebulizer with distilled water and let it dry completely.
11. Reinstall the components.
12. Check aerosol generation and optimize the gas flow of the nebulizer; see
page 50.
Cleaning of the nebulizer is now complete.
Summary of Contents for 76004567
Page 2: ......