Safety
— 12
SPECTRO ARCOS FHX3X — 08.02.2022
Warning! Optical radiation!
Plasma emits UV radiation. A protective pane is used to shield from radiation.
Operate the instrument only with the protective pane in place and the door
closed.
Warning of non-ionizing radiation!
A high-frequency emitter is used to operate the ICP, dependent on the type of
model. The electromagnetic emission is below the legally specified limit values.
Employees must be alerted that an electromagnetic emitter is used at the
workplace, and a cardiac pacemaker symbol must be installed!
Persons wearing a pacemaker or implanted defibrillator must not be
permitted in the work zone!
Warning! High voltage!
Higher electrical voltages occur inside the spectrometer.
Shut down the instrument immediately if it is damaged.
Never tamper with the safety devices! If you suspect that a safety device
does not work properly, shut down the instrument immediately!
Information
Compliance with the applicable occupational health and safety regulations at
the workplace and the samples used (hazardous substances) is mandatory.
Compliance with the safety data sheets of the hazardous substances is
mandatory
Information
In order to ensure the availability and operational safety of the equipment, we
recommend having the system checked for correct operation and performance at
least once a year by specialized technicians. Addresses of service organizations
authorized by SPECTRO can be found at www.spectro.com.