PN 9000-246-VE, Rev D
Setting the Units
The
UNITS
configuration screen allows you to select the Unit of measurement
in which the analyte will be displayed. Certain analytes or measurements are
limited to related units; therefore some units cannot be used for some ana-
lytes.
The Unit of measurement is factory-set according the sample stream
composition provided by the customer. Changing the Unit of measure-
ment can result in erroneous measurement outputs. Some analytes
cannot be measured using certain units. Consult with your AMETEK
representative or AMETEK Service personnel before changing the Unit.
Figure 3-8.
Unit Settings menu.
UNITS
( ppmV )
lb/mmscf
To change the measurement unit:
1. From the
CONFIG
menu, use the
Up
or
Down
arrow keys to select
Units
.
Press
Enter
to view the
Units
sub-menu.
2. Use the
Up/Down
arrow keys to view the list of Units (
PPMV
,
Lb/MMscf
,
mg/Nm3
,
dewC
,
dewF
,
ppmW
, and
Percent
).
Important information about changing Units:
• Moisture measurement units include
PPMV
,
Lb/MMscf
,
mg/Nm3
,
DewpointC, DewpointF
.
• Concentration measurement units include
PPMV
,
Lb/MMscf
,
mg/Nm3, DewpointC, DewpointF, ppmW, %
.
• Dew Point measurement units include
dewC
or
dewF
.
• Dew Point Calculations require a Process Pressure Value.
- For Dew Point in Centigrade, enter process pressure values in units
of BAR absolute.
- For Dew Point in Fahrenheit, enter process pressure values in units
of PSIA.
User Interface/5100P Application Software
|
3-11
Summary of Contents for 5100P
Page 1: ...5100P TDLAS Analyzer USER MANUAL PN 9000 246 VE Rev D 5000 016 PI REV A...
Page 10: ...PN 9000 246 VE Rev D This page intentionally left blank x 5100P TDLAS Analyzer...
Page 28: ...PN 9000 246 VE Rev D This page intentionally left blank 2 14 5100P TDLAS Analyzer...
Page 64: ...PN 9000 246 VE Rev D This page intentionally left blank 3 36 5100P TDLAS Analyzer...
Page 82: ...PN 9000 246 VE Rev D This page intentionally left blank 5 4 5100P TDLAS Analyzer...