Safely Powering Down the 5100P Analyzer to Perform
Maintenance
For certain maintenance procedures, the analyzer must be powered down.
Before powering down the analyzer read and follow the Warnings and
Cautions under “Safety Considerations,” earlier in this chapter, and
under “Personnel and Equipment Safety Information” following the
Table of Contents near the beginning of this manual.
Before proceeding, test the area around the analyzer for flammable
gases. If an explosive gas atmosphere is present, do not disconnect
power from the analyzer or any alternate power sources that supply
power to the analyzer components. Proceed only when the area is
found to be safe.
Before performing maintenance on the 5100P Analyzer, power down the ana-
lyzer by pressing the
Power On/Off
button on the front panel.
If process gas and vent lines are connected, disconnect the lines and replace
the caps on the Sample Panel connections to prevent unwanted ambient
items like dust and water from entering the analyzer sample path.
4-6 | 5100P TDLAS Analyzer