PN 9000-246-VE, Rev D
Low Pressure Sample Panel (Optional Operation)
The Low Pressure 5100P Sample Conditioning Panel (Figure 2-2.2) operates
with a raw Sample gas pressure of 0.06–1.03 BARG (1–15 PSIG).
Upstream of the AMETEK Sample Panel Inlet Valve, a pressure regulator (in-
stalled by customer/end user) is used to reduce the Sample gas pressure to the
recommended operating pressure of the 5100P Gas Analyzer (0.4–0.7 BARG
[6–10 PSIG]).
The low pressure Sample gas is passed through the Filter (membrane separa-
tor) to remove any entrained (trapped) liquids and particulate matter before
the gas is passed through the Flow Meter.
Figure 2-2.2.
Sample Panel Flow Diagram,
Maximum 15 PSIG (Low
Pressure version).
FI
SAMPLE PANEL-NON REGULATED
TOP VIEW
INLET VALVE
GENIE FILTER
DRAIN VALVE
FLOWMETER
WITH VALVE
OUTLET FITTING
(TO INSTRUMENT)
GAUGE
MAX PANEL INLET PRESSURE: 15 PSI
INSTRUMENT INLET/PANEL OUTLET: 101 KPA +/- 2.5 KPA
FLOW RATE: 1-2 SLPM
INLET VALVE
GAUGE
DRAIN
VALVE
DRAIN
SAMPLE PANEL
FLOW SCHEMATIC
5100P
ANALYZER
FLOWMETER
WITH VALVE
6–10 PSIG
15 PSIG
Connecting and Starting Up the 5100P
|
2-7
Summary of Contents for 5100P
Page 1: ...5100P TDLAS Analyzer USER MANUAL PN 9000 246 VE Rev D 5000 016 PI REV A...
Page 10: ...PN 9000 246 VE Rev D This page intentionally left blank x 5100P TDLAS Analyzer...
Page 28: ...PN 9000 246 VE Rev D This page intentionally left blank 2 14 5100P TDLAS Analyzer...
Page 64: ...PN 9000 246 VE Rev D This page intentionally left blank 3 36 5100P TDLAS Analyzer...
Page 82: ...PN 9000 246 VE Rev D This page intentionally left blank 5 4 5100P TDLAS Analyzer...