6-16 | Model 5100 HD Analyzer
40
111
30
“Cell pres Hi”
Warning
254, 495, 496
Cell pressure (254) is outside of a band
set by min/max pressure (495, 496)
111
31
“Cell pres Low”
Warning
111
32 "Span error"
Warning
Error in span calibration
112
1 "Zero error"
Warning
Error in zero calibration
113
2 "User error"
Warning
User error detected from the user analog
input
Common Modbus Registers
ID
Access Data
Type
mnemonic
MB
Starting
Regs
MB Reg
Size
Description
6
rs
char[20]
SERIAL
0
10 Serial Number
7
rs
char[6]
MAC
10
3
MAC address
8
rws
uint
IP
13
2
IP Address
9
rws
uint
SUBNET
15
2 IP Subnet Mask
10
rws
uint
GW
17
2 Gateway
11
rws
uint
DNS
19
2 DNS
12
rws
uint16
DHCPE
21
1 DHCP enable. 1=disabled, 2=enabled
13
rws
uint16
TCPE
22
1
TCP Enable. 0=disabled, 1=enabled
14
rws
uint
SVRIP
23
2 Server IP
100
rwf
int16
CFGFILEVER
25
1 Version of the params.xml file generated by
this
101
r
char[20]
VERSION
26
10 Software Version
102
r
char[20]
NAME
36
10 Model Name
103
rw
uint
IU1BAUD
46
2
Internal UART1baud rate (MODBUS)
104
rwf
uint
IU2BAUD
48
2
Internal UART2 baud rate (DISPLAY)
105
rwf
uint
EU1BAUD
50
2
External UART1 baud rate (RS232)
106
rwf
uint
EU2BAUD
52
2
External UART2 baud rate
107
r
uint16
FWVERHRB
54
1 Interface Board FW version
108
r
uint16
CARDID
55
1 0x20: HC121, 0x40: HC122
109
rw
int16
MBPARITY
56
1 Parity. 0=none, 1=even, 2=odd
110
rwf
int16
MBSTOPBIT
57
1 Stop bits. 1=1 stop bits, 2=2 stop bits
111
rw
uint16
MBUSADD
58
1
Modbus Address
112
rwf
uint16
MBUSPORT
59
1
Modbus Port. 0=Disable, 1=Enable
113
rwf
uint
MBUSTO
60
2
Modbus Timeout (ms)
114
r
char[20]
SYSTIME
62
10 System Time
115
r
char[20]
SYSDATE
72
10 System Date
117
rw
char[20]
CONCUNIT
83
10 Concentration Unit of Measure for analyte 1
120
rw
char[10]
PRESUNIT
95
5 Pressure Unit of Measure, case sensitive.
Valid strings include psi, psia, inH20, inHg,
bar, cmH20, kg/cm2, kPa, mbar, micron,
mmHg, oz/in2, torr, atm
121
rw
char[10]
TEMPUNIT
100
5 Temp Unit of Measure, case sensistive. Valid
strings include degreeC, degreeF, degreeK
122
r
uint
DATAVALID
105
2 Data Valid. 0=Not Valid, 1=Valid
123
r
int16
DMODE
107
1
0=Line-
lock mode, 1=Sample, 2=Calib
125
r
uint
SYSALARM1
109
2 Bit
-
mapped system alarms
126
r
uint
SYSALARM2
111
2 Bit
-
mapped system alarms
127
r
uint
USERALARM
113
2 Bit
-
mapped system alarms
128
r
uint
CONCALARM
115
2 Bit mapped concentration alarms
Common Modbus Registers
41
129
r
int16
SYSSTATUS
117
1 System Status.
0=Everything ok, Not
0=something is wrong
130
r
int16
SYSSTATE
118
1
System State. 100=Not Ready, 201=Measure
Run, 502=Idle, 600=diag
133
r
float
CELLTEMPD
123
2 Cell temp for display
134
r
float
ANAOUT1
125
2
Current Output 1
135
rw
float
ANAOUT1O
127
2
Current Output 1 Offset
136
rw
uint16
ANAOUT1P
129
1
Current Output 1 Parameter
137
rw
float
ANAOUT1S
130
2
Current Output 1 Span
138
rw
int16
ANAOUT1T
132
1
Current Output 1 Output Type. 0=4-
20 mode,
1=0
-
20 mode.
If overange enable (ANAOUT1OR = 1)
0 = 3.78
-
20.5, 1 = 0
-
20.5, high or low
values can also be set by fault value, See
Fault option (ANAOUT1F)
139
rw
uint16
ANAOUT1OR
133
1
Current Output 1 Over range Enable.
0=disabled, 1=enabled
140
rw
uint16
ANAOUT1F
134
1
Current Output 1 Fault Option. 0=No fault,
1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-
20 mode
(ANAOUT1T = 0):
Low end of range becomes 3 mA
Output ratio bec
omes 0.
141
r
float
ANAOUT2
135
2
Current Output 2 -
142
rw
float
ANAOUT2O
137
2
Current Output 2 Offset -
143
rw
uint16
ANAOUT2P
139
1
Current Output 2 Parameter-
144
rw
float
ANAOUT2S
140
2
Current Output 2 Span -
145
rw
int16
ANAOUT2T
142
1
Current Output 2 Output Type. 0=4-
20 mode,
1=0
-
20 mode.
If overange enable (ANAOUT2OR = 1)
0 = 3.78
-
20.5, 1 = 0
-
20.5, high or low
values can also be set by fault value, See
Fault option (ANAOUT2F) -
146
rw
uint16
ANAOUT2OR
143
1
Current Output 2 Over range Enable.
0=disabled, 1=enabled
147
rw
uint16
ANAOUT2F
144
1
Current Output 2 Fault Option. 0=No fault,
1=Low fault, 2=High fault.
If high fault (2):
High end of range becomes 22 mA
Output ratio becomes 1
If low fault (1) AND in 4-
20 mode
(ANAOUT2T = 0):
Low end of range becomes 3 mA
Output ratio becomes 0.
148
r
uint16
RELAY1
145
1
Relay 1. 0=OPEN, 1=CLOSED
149
rw
uint16
RELAY1D
146
1
Relay 1 Delay. Holds the relay output for this
set amount of time.
150
rw
uint16
RELAY1F
147
1
Relay 1 Function. 0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
151
rw
uint16
RELAY1S
148
1
Relay 1 Source. The source parameter
number used to evaluated the relay function.
152
rw
float
RELAY1V
149
2
Relay 1 Value. The value used to
evaluate
the relay function.
153
r
uint16
RELAY2
151
1
Relay 2. 0=OPEN, 1=CLOSED
Summary of Contents for 5100 HD
Page 23: ...Overview 1 11 Space Coupled Option Schematic...
Page 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Page 39: ...Installation Start Up 3 11 I O Signals...
Page 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Page 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Page 120: ...6 20 Model 5100 HD Analyzer...