3-18 | Model 5100 HD Analyzer
Ethernet Cable Connections
When using the Ethernet interface on the 5100 HD, the minimum requirement for
the Ethernet cable is a CAT 5e (maximum 100 m) cable to connect the analyzer
to the network. The maximum distance between the analyzer and the hub is
100 m. For outdoor analyzer installations, hazardous locations, and/or lengths
greater than 100 m, special requirements may be necessary. Contact your network
administrator for more information.
To install the Ethernet cable, follow the steps below:
1. Disconnect the power from the analyzer and remove the cover on the
customer connection box.
2. Route the Ethernet cable through the cable entry port and into the box.
3. Plug the cable into the LAN RJ45 connector.
4. Route the cable to the network connection in the plant and connect it.