4-14 | Model 5100 HD Analyzer
Background Gas
This selection allows for the proper background gas to be set. It is critical that
the background is chosen to represent the actual process gas as closely as pos-
sible, The background gas is typically chosen at the factory during calibration
to ensure the best accuracy, based on the customer application data. However, it
may be changed using this utility. Changing the background gas results in the ap-
plication of a multiplication factor to the factory calibration curve
Units
Use the
Unit
function to select the unit of measurement for the analyzer. (The
map below is shown as an example using moisture.)
MOISTURE
UNITS
Dewpoint F
Dewpoint C
PPMV
LB/MMSCF
MG/NM3
PPMV
LB/MMSCF
MG/NM3
PPMV
LB/MMSCF
MG/NM3
*
High Pressure
Pressure
*
Low Pressure
*
Fixed Pressure
Transmitter
Secondary Unit
Pressure
*
Fixed Pressure
Transmitter
High Pressure
*
Low Pressure*
Secondary Unit
Menu for the Unit key.
Select the unit for the readings from the list. Press
Enter
.
If you choose either
Dew Point C
(dewC) or
Dew Point F
(dewF), you must also
select one of the following:
Fixed Pressure Enter the value in units of kPa
Transmitter
Enter
High Pressure
and a
Low Pressure
in units of kPa
Secondary Unit Choose moisture unit: ppmV, lb/mmscf, or mg/nm
3
If you change the units, remember to make any necessary changes to
the Alarm and Analog Range settings.
NOTE
Summary of Contents for 5100 HD
Page 23: ...Overview 1 11 Space Coupled Option Schematic...
Page 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Page 39: ...Installation Start Up 3 11 I O Signals...
Page 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Page 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Page 120: ...6 20 Model 5100 HD Analyzer...