6-12 | Model 5100 HD Analyzer
35
335
rwf
float
RLEVEL10
467
2
Response value level 7
336
rwf
float
OLEVEL10
469
2
Output level 7
337
rwf
float
ZVALUE
471
2 Power correction Z value
338
rw
float
ZVALUE2
473
2 Power correction Z value 2
339
rwf
float
PRESSCOEFF1_2
475
2 Pressure compensation coefficient 1
340
rw
float
PRESSCOEFF2_2
477
2 Pressure compensation coefficient 2
341
rw
float
PRESSCOEFF3_2
479
2 Pressure compensation coefficient 3
342
rw
float
PRESSCOEFF4_2
481
2 Pressure compensation coefficient 4
343
rw
int16
TECMIDP_2
483
1
Midpoint of the DAC range for TEC
control
344
rwf
float
REFCONC
484
2
Concentration of Reference call
345
rwf
float
CELLDT
486
2 Heater Deriviative Time
346
r
float
CELLHDC
488
2 Heater Duty Cycle
347
rwf
float
CELLIT
490
2 Heater Integral Time
348
rwf
float
CELLPB
492
2 Heater Proportional Band
349
rwf
float
CELLST
494
2 Cell Sample Time
350
rwf
float
REFDT
496
2 Heater Deriviative Time
351
r
float
REFHDC
498
2 Heater Duty Cycle
352
rwf
float
REFIT
500
2 Heater Integral Time
353
rwf
float
REFPB
502
2 Heater Proportional Band
354
rwf
float
REFSP
504
2 Heater Set Point
355
rwf
float
REFST
506
2
Ref Sample Time
356
rwf
int16
TECMIDP
508
1
Midpoint of the DAC range for TEC
control
357
rwf
float
PRESSCOEFF1
509
2 Pressure compensation coefficient 1
358
rwf
float
PRESSCOEFF2
511
2 Pressure compensation coefficient 2
359
rwf
float
PRESSCOEFF3
513
2 Pressure compensation coefficient 3
360
rwf
float
PRESSCOEFF4
515
2 Pressure compensation coefficient 4
361
r
float
ALYTE1CONS
517
2
Analyte 1 Concentration in other unit
362
rw
char[20]
CONCUNITS
519
10 2nd Concentration Unit of Measure
for analyte 1
363
r
uint
DATAVALID_2
529
2 Data Valid. 0=Not Valid, 1=Valid
364
r
int16
DMODE_2
531
1
0=Line-
lock mode, 1=Sample,
2=Calib
365
rw
int16
COMPPT_2
532
1 Pressure and temp compensation.
0=none, 1=pressure, 2=temperature,
3=both
366
rw
float
SEALPRESS
533
2 Process seal pressure span
367
rw
float
SEALPRESOS
535
2 Process seal pressure offset
368
rw
float
SEALPRESIIR
537
2
Process seal pressure IIR
369
rw
float
SEALPRESMAX
539
2 Process seal max
370
rw
float
SEALPRESMIN
541
2 Process seal min
371
rw
float
LASTHS
543
2 Sample channel's threshold
372
rw
float
LASTHS2
545
2 Sample channel 2's threshold
373
rw
uint
DUALCONC
547
2 0=single, 1=dual concentration from
one spectrum
374
r
float
TECTEMP_2
549
2
TEC Temperature Value
375
rw
int16
TECAT_2
551
1
Heater Actuator Type. 0=Reverse,
1=Forward.
376
rw
float
TECDEVT_2
552
2 Heater Deviation Tolerance. Ignored if
0. Otherwise high or low limit alarm
are set when value is out bounds of 1
deviation
377
rw
float
TECDT_2
554
2 Heater Deriviative Time
378
r
float
TECHDC_2
556
2 Heater Duty Cycle
379
rw
float
TECIT_2
558
2 Heater Integral Time
380
rw
float
TECPB_2
560
2 Heater Proportional Band
36
381
rw
float
TECSP_2
562
2 Heater Set Point
382
rw
float
TECST_2
564
2
TEC Sample Time
383
rw
float
TECHYS_2
566
2
TEC hysterisi
384
rw
int16
STATICPT_2
568
1
Pressure (STCELLP) and temp
(STCELLT) static value usage.
0=neither, 1=use static pressure,
2=use static temperature, 3= use both
385
rw
float
STCELLT_2
569
2 Static cell temperature. Use by
setting STATICPT to 2 or 3
386
rw
float
STCELLP_2
571
2 Static cell pressure. Use by setting
STATICPT to 1 or 3
387
rw
float
REFREFCONC_2
573
2
Reference soncentration of the
reference cell at
388
rw
float
TECIIR_2
575
2
TEC IIR Filter Weight
389
rw
int16
WLTOL1_2
577
1 Wavelength Tolerance 1
390
rw
int16
WLTOL2_2
578
1 Wavelength Tolerance 2
391
rw
float
IOFFSET_2
579
2 Injection current offset
392
rw
float
LSTWLRES_2
581
2 Temperature Wavelength resolution
393
rw
int16
MEDSIZE_2
583
1 Median filter size
394
r
int16
CTRINDX_2
584
1 Spectrum center index
395
rwf
float
TEMPREF_2
585
2
Reference temperature for this diode
396
rw
int16
CONCSTART_2
587
1 Starting point of concentration window
397
rw
int16
CONCEND_2
588
1
End point
of concentration window
398
rw
int16
CONCPEAK_2
589
1 Peak index for linelock
399
rw
float
RLEVEL1_2
590
2
Response value level 1
400
rw
float
OLEVEL1_2
592
2
Output level 1
401
rw
float
RLEVEL2_2
594
2
Response value level 2
402
rw
float
OLEVEL2_2
596
2
Output level 2
403
rw
float
RLEVEL3_2
598
2
Response value level 3
404
rw
float
OLEVEL3_2
600
2
Output level 3
405
rw
float
RLEVEL4_2
602
2
Response value level 4
406
rw
float
OLEVEL4_2
604
2
Output level 4
407
rw
float
RLEVEL5_2
606
2
Response value level 5
408
rw
float
OLEVEL5_2
608
2
Output level 5
409
rw
float
RLEVEL6_2
610
2
Response value level 6
410
rw
float
OLEVEL6_2
612
2
Output level 6
411
rw
float
RLEVEL7_2
614
2
Response value level 7
412
rw
float
OLEVEL7_2
616
2
Output level 7
413
rw
float
RLEVEL8_2
618
2
Response value level 7
414
rw
float
OLEVEL8_2
620
2
Output level 7
415
rw
float
RLEVEL9_2
622
2
Response value level 7
416
rw
float
OLEVEL9_2
624
2
Output level 7
417
rw
float
RLEVEL10_2
626
2
Response value level 7
418
rw
float
OLEVEL10_2
628
2
Output level 7
419
rw
float
ALPHA2
630
2
Alpha value for wir filter
420
rw
float
BETA2
632
2 Beta value for wir filter
421
rw
float
PWRALPHA_2
634
2 Power correction alpha
422
rw
int16
AVESIZE_2
636
1 Beta value for wir filter
423
rw
float
ZVALUE_2
637
2 Power correction Z value
424
rw
float
SAMPHDC_2
639
2 Duty cycle for the sample heater
425
rw
float
ZVALUE2_2
641
2 Power correction Z value 2
426
r
float
LASMPPWR
643
2
Laser power of
sample channel
427
rw
float
LASMPREF
645
2
Laser power reference of sample
channel
428
r
float
LASMPPWR_2
647
2
Laser power of sample channel 2
Summary of Contents for 5100 HD
Page 23: ...Overview 1 11 Space Coupled Option Schematic...
Page 24: ...1 12 Model 5100 HD Analyzer This page intentionally left blank...
Page 39: ...Installation Start Up 3 11 I O Signals...
Page 50: ...3 22 Model 5100 HD Analyzer This page intentionally left blank...
Page 100: ...5 16 Model 5100 HD Analyzer This page intentionally left blank...
Page 120: ...6 20 Model 5100 HD Analyzer...