User Guide
6 - 4
Particulate Measurement System
4750-PM
6.5 Troubleshooting
Refer to the Section 6.1 Safety Information section provided at the start of
this chapter and as given in the sections referenced below.
6.5.1 Common Issues
Symptom
Solution
Further options
Power/Comms
LED is OFF
Check power cables and fuse.
If necessary, replace the fuse
(see Section 6.4).
Alarm LED is
flashing RED
or AMBER
Likely cause: the sensor reading
is above the alarm level (this
may also be due to a sensor
check fail, see below).
Contact AMETEK
Land or our local
representative for
assistance.
Sensor check
has failed
(1) Rerun the sensor tests to
determine which check is
causing the alarm.
(2) Carry out the sensor
cleaning procedure (see
Section 6.3), then rerun
the checks.
(3) If the sensor checks fail
repeatedly, the sensor PCB
may need replacing
6.5.1 Alarms
The
4750-PM
monitors threshold conditions that can be set in the
instrument to comply with local regulations and guidelines. Alarm conditions
are indicated by an alarm LED (refer to Fig. 3-2).
Alarm
Description
Zero fault
Fault with the sensor electronics.
Please contact AMETEK Land or our local
representative for assistance.
Span fault
Fault with the sensor electronics or the laser (see
troubleshooting sensor check failures above).
Please contact AMETEK Land or our local
representative for assistance.
Air Purge Flow Failure
> Low Flow Alarm
The sensor requires a continuous supply of purge
air and may have become contaminated.
The sensor optics should be cleaned (see Section
6.3).
Summary of Contents for 4750-PM
Page 4: ...Particulate Measurement System User Guide 4750 PM Blank...
Page 6: ...INTRODUCTION 1...
Page 11: ...TECHNICAL DATA 2...
Page 15: ...2 3 User Guide Particulate Measurement System 4750 PM...
Page 18: ...PRODUCT DESCRIPTION 3...
Page 24: ...INSTALLATION 4...
Page 36: ...OPERATION 5...
Page 43: ...5 7 User Guide Particulate Measurement System 4750 PM Fig 5 2 Engineering Menu...
Page 53: ...MAINTENANCE 6...
Page 59: ...RESETTING THE SENSOR 7...
Page 61: ...GENERAL NOTICES 8...
Page 63: ...4750 PM Particulate Measurement System User Guide...