3 - 1
User Guide
Particulate Measurement System
4750-PM
3
Product Description
3.1 About the 4750-PM
The
4750-PM
backscatter sensor with two external status indicators and
protected, internal user controls is suitable for measuring dust concentrations
and monitoring filter performance in medium to large stacks or ducts
for industrial and combustion processes. The sensor can measure dust
concentrations as low as 0.1 mg/m³, making it particularly suitable for the
low dust levels found after bag filters. It is unaffected by changes in flue gas
velocity associated with variable speed fans or plant-load changes and, for a
given particle size and type, the instrument response is proportional to dust
concentration.
The instrument has been designed to comply with the TUV suitability testing
scheme to EN 15267 and, due to its wide dynamic operating range, may be
used in higher dust loading applications as well as for process optimization
applications.
3.2 Components
3.2.1 4750-PM Sensor
1 Sensor enclosure with hinged lid
5 Purge Air Flow sensor
2 Hinged sensor body
6
Purge air supply (flexible tube with
hose clip)
3
Inbuilt flange (sensor-to-stack
connection)
7
M20 cable glands (3 off)
4 Quick-release clamp
8 Sensor interface with external status
indicators (2 off)
Fig. 3-1 4750-PM Sensor
Summary of Contents for 4750-PM
Page 4: ...Particulate Measurement System User Guide 4750 PM Blank...
Page 6: ...INTRODUCTION 1...
Page 11: ...TECHNICAL DATA 2...
Page 15: ...2 3 User Guide Particulate Measurement System 4750 PM...
Page 18: ...PRODUCT DESCRIPTION 3...
Page 24: ...INSTALLATION 4...
Page 36: ...OPERATION 5...
Page 43: ...5 7 User Guide Particulate Measurement System 4750 PM Fig 5 2 Engineering Menu...
Page 53: ...MAINTENANCE 6...
Page 59: ...RESETTING THE SENSOR 7...
Page 61: ...GENERAL NOTICES 8...
Page 63: ...4750 PM Particulate Measurement System User Guide...