User Guide
5 - 8
Particulate Measurement System
4750-PM
5.3.2 Engineering Functions [ENG]
Several test, reset and 4-20 mA adjustment functions can be accessed via
this menu.
Setting
Description
RSET
Resets the sensor to factory (=default) values.
420Z
Trims 4-20mA Zero Offset (4mA).
420F
Trims 4-20mA Full Scale (20mA).
T.REL
Tests the relays.
T.DIG
Tests the digital input.
T.LED
Tests the LEDs.
Trimming the 4-20mA Outputs [420Z, 420F]
1) Scroll DOWN to the ENG menu.
2) Press RIGHT repeatedly to access the
420Z
and
420F
submenus. To
adjust the values, proceed as follows:
a. When accessing the 4-20 mA Zero [
420Z
] setting, the sensor
nominally outputs 4 mA. Use the UP/DOWN buttons to adjust the
4-20 mA Zero calibration value (default: 223) to trim the 4 mA
value.
b. Similarly, when accessing the 4-20 mA Full Scale [
420F
] setting,
the sensor nominally outputs 20 mA. Use the UP/DOWN buttons to
adjust the 4-20 mA Full Scale calibration value (default: 4385) to
trim the 20 mA value.
Test Functions [T.REL, T.DIG, T.LED]
1) Scroll DOWN to the
ENG
menu.
2) Press RIGHT repeatedly to access the test function submenus.
3) Use the UP/DOWN buttons to change the values. The table below shows
the operation of the test functions.
Setting
Value
Meaning
Relays [T.REL]
Set to 1
Set to 2
Relay 1 Open
Relay 2 Open
Digital Input
[T.DIG]
Open
Apply 24 V to close
Display 0000
Display 0001
LEDs [T.LED]
Set to 0
Set to 3
All LEDs =
GREEN
Self-check LED =
RED
Summary of Contents for 4750-PM
Page 4: ...Particulate Measurement System User Guide 4750 PM Blank...
Page 6: ...INTRODUCTION 1...
Page 11: ...TECHNICAL DATA 2...
Page 15: ...2 3 User Guide Particulate Measurement System 4750 PM...
Page 18: ...PRODUCT DESCRIPTION 3...
Page 24: ...INSTALLATION 4...
Page 36: ...OPERATION 5...
Page 43: ...5 7 User Guide Particulate Measurement System 4750 PM Fig 5 2 Engineering Menu...
Page 53: ...MAINTENANCE 6...
Page 59: ...RESETTING THE SENSOR 7...
Page 61: ...GENERAL NOTICES 8...
Page 63: ...4750 PM Particulate Measurement System User Guide...