65
098-0007 Rev B
Steps
Actions
cut-off current. Turn the screw until the
green LED on SSR1 goes out completely.
Step 8:
Stop the heat disinfection mode on the
MROC. Place the MROC in to normal
operation.
Step 9:
Verify that the heating element on the
Centurion Plus Heat disinfection device is
activated while the MROC is running in
normal operation.
NOTE:
Green LED on SSR1 will be
illuminated when heater is on.
Step 10:
With the current switch set, reinstall the
top cover and side covers onto the
Centurion Plus Heat disinfection device.
Be sure to manually shut off the heating
element using diagnostics menu page 1.
Step 11:
Navigate to the diagnostics menu page 2
and manually drain the tank on the
Centurion Plus.
Step 12:
Return the device to service.
Summary of Contents for 000-HS120
Page 2: ......
Page 24: ...22 098 0007 Rev B Right Side...
Page 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Page 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Page 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Page 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Page 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Page 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Page 69: ...67 098 0007 Rev B...
Page 70: ...68 098 0007 Rev B...
Page 71: ...69 098 0007 Rev B...
Page 72: ...70 098 0007 Rev B...
Page 73: ...71 098 0007 Rev B...
Page 74: ...72 098 0007 Rev B...