098-0007 Rev B
Steps
Actions
approximately 5 minutes for the heating
element to turn on.
Step 6:
Manually activate the heater on the
Centurion Plus using diagnostics page 1.
Verify that the red LED light on the
current switch is flashing rapidly and the
Green LED on SSR1 is off.
NOTE:
If the green LED on SSR1 flashes
rapidly, turn off the heating element to
avoid blowing the fuses on the Centurion
Plus.
Step 7:
If the Green LED on SSR1 flashes, or is
illuminated, use a small flat head
screwdriver to adjust the screw on the
current switch. Turn the screw clockwise
to lower the cut-off current or turn the
screw counter clockwise to increase the
LED will flash every 2
seconds when current draw
does not exceed set-point.
LED will flash every 0.5
seconds when set-point is
exceed (heating element
disabled)
Turn screw clockwise (CW)
to lower set-point. Turn
screw counter clockwise
(CCW) to increase set-point.
Green LED will be
illuminated when heater is
on. If LED is flashing, heater
is not drawing 1 kW. Light
should be out when heater is
disabled by current switch
Summary of Contents for 000-HS120
Page 2: ......
Page 24: ...22 098 0007 Rev B Right Side...
Page 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Page 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Page 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Page 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Page 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Page 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Page 69: ...67 098 0007 Rev B...
Page 70: ...68 098 0007 Rev B...
Page 71: ...69 098 0007 Rev B...
Page 72: ...70 098 0007 Rev B...
Page 73: ...71 098 0007 Rev B...
Page 74: ...72 098 0007 Rev B...