Check diagnostics page 2,
FS2 should be displayed as
“Open”
Active alarms preventing
operation
Verify that there are no
active alarms.
Water does not pass
through device in standby
Valve SV4 has failed closed
Shut power off to the
Centurion Plus Heat
disinfection device. Water
should pass through with
power off
If water does not pass
through device, disconnect
RO from Centurion Plus Heat
disinfection device to
continue patient treatment
Valve SV1 has failed in the “N.C”
position (diverts all water to
tank)
Check float switch status on
Diagnostics page 2.
Water does not drain from
tank during manual tank
drain mode
System pump failure
See “Pump does not turn on”
section
Valve SV2 failed in normal
position (closed to drain)
Verify there is water in tank
of system using diagnostics
menu
Manually activate pump and
valve SV2. Water should be
pumped out of the system
drain port when the feed
water RO is off
If water does not exit device
through drain port valve SV2
is not functioning properly
Manually activate valve SV3
and place the system into
tank drain mode as a tank
drain back up
Lower float switch fails open
causing false tank level display
Remove tank lid to verify the
water level in the tank
Hot water sent to drain
during dispensing mode
Valve SV2 failed open
Fill the tank on the Centurion
Plus with water
Shut off the feed water RO
Manually turn on the pump.
Verify that water does not
exit the system drain port
MROC break tank over flows
during dispensing mode
Valve SV3 fails in normal
condition (Open to RO return)
Fill tank on Centurion Plus
with RO water
Shut off MROC
Manually turn on valve SV3
using diagnostics menu
Manually turn on pump
Verify that water exits device
through system drain port
Summary of Contents for 000-HS120
Page 2: ......
Page 24: ...22 098 0007 Rev B Right Side...
Page 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Page 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Page 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Page 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Page 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Page 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Page 69: ...67 098 0007 Rev B...
Page 70: ...68 098 0007 Rev B...
Page 71: ...69 098 0007 Rev B...
Page 72: ...70 098 0007 Rev B...
Page 73: ...71 098 0007 Rev B...
Page 74: ...72 098 0007 Rev B...