APPENDIX C: CALIBRATION
HF28 DC RESISTANCE WELDING SYSTEM
C-2
990-372
Calibration Procedure
Initial Calibration Setup
1.
Connect the calibration setup to the
Control as shown.
2.
Turn the Control ON.
3.
In the
MAIN MENU
press
4
to enter the
calibration screen.
<MAIN MENU>
1. SETUP
6. RELAY
2. WELD COUNTERS
7. RESET DEFAULTS
3. COPY A SCHEDULE
8. CHAIN SCHEDULES
4.
CALIBRATION
5. SYSTEM SECURITY
Number Select an item
4.
Press
1
for
HF28 CALIBRATION
which
will bring up the
CAUTION
screen on
the right. Press the
arrow.
CAUTION
CALIBRATION SHOULD BE PERFORMED BY A
QUALIFIED TECHNICIAN ONLY.
REFER TO MANUAL FOR CALIBRATION SETUP.
next, MENU menu
5.
Press
2
to calibrate the HF28.
<PRE-CALIBRATION>
1. TEST HF28 (T-232 REQUIRED)
2. CALIBRATE HF28
3. RESET CALIBRATION
4. SET CURRENT SHUNT VALUE
5. TEST SERIAL PORT
Menu menu