Fluo-100 Portable Fluorometer Operation Manual
Chapter 5
Software Introduction
17
Detecting datas will be saved automatically after detecting. You can delete the useless data.
It can save 30 datas of each sample detecting at most. Once it is full, the earliest data will be
covered. The data can be checked in “Report”.
5.2.2.3 Calibration
You can calibrate the instrument via using a small amount of samples (1-3) to improve the
accuracy of detection (Fig 5.21). It also helps to save time and samples of setting standard
curve.
Fig5.21 Calibration Interface
NOTE:
You can prepare 1-3 samples to calibrate the instrument according to the below
principles:
1. Single point calibration, using blank solution to calibrate.
2. Two point calibration, using a high and a low concentration sample to calibrate.
3. Three point calibration, using a high,middle and a low concentration sample to calibrate.
Normally, two point calibration is enough.
5.3 Report
Click “Report” to enter the report interface (Fig5.22).