Advanced parameters and commands
Goldeye Technical Manual
V2.2.0
66
Example
Activate the two-point correction
>
E=1[CR]
The parameters
A
and
B
define both files out of which the reference images
for the two-point correction data are loaded. If various correction data sets are
available, the reference images in the file system are stored each in a closed
sequence of file numbers. The loading process following the camera start
begins at the first cold (low) reference image
A
. Then it is continued with the
accompanying warm (high) reference image
B
. In the following step it is
attempted to load the file numbers
A+1
, then
B+1
and so on. This process
is continued as long as one file number is missing within the file system or the
maximum number of correction sets (actually 32) had been achieved.
In case during operation the value of
A
or
B
is new adjusted, this change
causes the immediate data load out of this file. Different to the automatic load-
ing procedure following the switch-on of the camera, the data are exclusively
copied within the memory page
S
Select a correction data set (S)
Figure 22: Schematic process of two-point correction
Load process
Cold reference image
A
Cold reference set value
J
Correction
data
0x ??????
...
...
Correction
data
0x ??????
0
0
n
n
Warm reference image
B
Warm reference set value
K
Correction
data
0x ??????
...
...
Correction
data
0x ??????
0
0
n
n