Basic parameters and commands
Goldeye Technical Manual
V2.2.0
61
Select a correction data set (S)
All
Goldeye xy-...
cameras are equipped with multiple correction data sets for
different operation conditions. The image quality can be improved by activating
another correction data set if the environmental conditions have changed.
For a quick access to the different correction data sets, the correction data
memory is subdivided into 32 single pages. Each of these pages can store a com-
plete set of correction data (parameters
A
and
B
). At all times only one page
can be active simultaneously and all eventual modification of the correction
data takes place within this page. For each correction memory page a further
pair of registers for the parameters
J
and
K
is available. In order to enable
an access on the values
J
and
K
of each correction data set, the page address
also serves as access address to the register pairs.
The memory pages are filled with data starting at the address 0. The highest
valid address depends on the number of correction data sets stored in the cam-
era, however it can not exceed 31 (0x1F). If e.g. 27 correction data sets are
available this would result in a maximum valid page address of 0x1A.
Both parameters,
A
and
B
together, affect the choice and the number of data
sets to be loaded.
By means of the parameter
S
the address of the active page can be adjusted:
Example
Activate the eleventh correction data set (address 10 = 0xA)
>
S=A[CR]
GigE feature name (CameraSpecialFeatures)
Feature Visibility Description
CorrectionDataSet
Beginner
Number of the correction data
set to activate.
(S=<value> command)
Table 28: GigE feature: CorrectionDataSet