Basic parameters and commands
Goldeye Technical Manual
V2.2.0
60
Example
Start the correction data selection once and activate the best data set based on
the internal statistics value.
>
k=0[CR]
S=0A
>
Goldeye xy-008
If
k
is not set to 0 the automatic search starts at approx. each
k
* 256
images. At a frame rate of e.g. 50 images per second the chronological resolu-
tion of the
k
value is approx. 5.1 seconds.
Goldeye xy-032
If
k
is not set to 0 the automatic search starts at approx. each
k
* 256
images. At a frame rate of e.g. 30 images per second the chronological resolu-
tion of the
k
value is approx. 8.53 seconds.
Example
Goldeye xy-008
The automatic correction data selection shall always be started approx. every
1200 seconds (20 minutes): 1200 sec. / 5.1 sec. per count ≈ 235 = 0xEB.
Goldeye xy-032
The automatic correction data selection shall always be started approx. every
1200 seconds (20 minutes): 1200 sec. / 8.53 sec. per count ≈ 140 = 0x8C.
>
k=EB [CR]
(example Goldeye xy-008)
S=0A
(...at once)
>S=0B
(...after each 235 * 256 images...)
(...)
Note
For more information on the corresponding commands see
Chapter
GigE feature name (CameraSpecialFeatures)
Feature Visibility Description
AutoCalibrateOnce
Beginner
Start the automatic calibration
once.
(k=0 command)
The processing of this com-
mand can take several seconds,
depending on the current
frame rate and the number of
correction data sets available.
Table 27: GigE feature: AutoCalibrateOnce