BDDF00144 Issue 1
26/40
Alaris™ GW 800 Volumetric Pumps
Flow Sensor Operation
Flow Sensor Operation
w
The flow sensor automatically monitors the infusion flow rate through the drip chamber. The flow sensor will cause
the Pump to alarm if a significant deviation from the set infusion rate occurs. The flow sensor will also be able to
detect empty containers. For this reason, use of a flow sensor is recommended when using an infusion set without
an anti-siphon valve.
Flow Sensor Usage
Infusion set with anti-siphon valve ?
Use flow sensor ?
YES
Optional
NO
Recommended
Model 180 Flow Sensor
Handles
Drip chamber
Flanges
Pull back handles
Flow sensor
interface
Flow
Sensor
Interface
Cover
1 . Plug the flow sensor into the flow sensor interface located on the top rear part of the Pump .
2 . Attach the Model 180 Flow Sensor to the drip chamber of the Infusion set, by pulling back the handles . Refer to the illustration
above .
3 . Proceed with load, priming, and set-up instructions as described in section 'Getting Started' .
NOTE:
Ensure drip chamber is half full and upright .
w
Always attach the flow sensor before starting an infusion.
Avoid using the flow sensor in direct sunlight.
Always ensure lens is clean.