BDDF00144 Issue 1
23/40
Alaris™ GW 800 Volumetric Pumps
Configurable Options
Configurable Options
w
The default settings are configurable as displayed in brackets in the table below. Each of the configurable options
has a code which must only be altered by Qualified Service Personnel with reference to the technical service manual
(TSM) for this product (Technical Service Manual reference: BDTM00005).
Description
Range
Default
Setting
Enable VTBI / time infusions
(On / OFF)
OFF
Maximum priming volume
(OFF, 1 to 40 ml)
40ml
Clear infusion parameters to zero on
power - up
(On / OFF)
OFF
Maximum VTBI in MICRO Mode
(0 .1 to 999 ml)
999ml
Bolus rate
(1 to 999ml/h)
400ml/h
Maximum bolus volume
(OFF, 1 to 99ml)
5ml
Keep vein open rate
(OFF, 1 .0 to 5 .0 ml/h)
5 .0ml/h
Air in line alarm volume - single bubble
(50,100, 250, 500µL)
100µl
Enable secondary infusion capability
(On / OFF)
OFF
Default occlusion pressure on power - up (Lo (250mmHg),Nor (350mmHg),
HI (500mmHg))
HI
Alarm volume level
(1 - 7)
4
Enable Micro mode
(On / OFF)
OFF
Maximum infusion rate
(1 - 999 ml/h)
999ml/h
Enable ASCII mode for communications
(On / OFF)
OFF
Enable Odd parity for communications
(On / OFF)
OFF
Set Pump address for communications
(1 - 250)
1
Flow Sensor Connection Mode**
(AUTO/On)
AUTO
Set - up of current time and date
(00:00 to 23:59)
(01/01/00 to 31/12/99)
N/A
Language selection
(EnGL, FrAn, dEut, ItAL, ESPA, SE,
nEd)*
EnGL
IrDA Communications Selection
(On / OFF)
On
Nurse Call Activation High Enabled
(On / OFF)
On
Drops per ml of fluid
(1 to 200)
20
Silent Mode
(On / OFF)
OFF
User select mode options
Pressure limit Enabled
Alarm volume Enabled
Timed infusions Enabled
Micro infusions Enabled
(On / OFF)
(On / OFF)
(On / OFF)
(On / OFF)
OFF
OFF
OFF
OFF
Flow sensor sensitivity level
(Nor, Hi)
Nor
*EnGL - English, FrAn - French, dEut - German, ItAL - Italian, ESPA - Spanish, SE - Swedish, nEd - Dutch.
** If infusion sets without anti-siphon valves are being used, it is recommended to change the flow sensor connection mode setting to
On
. With
On
selected, the Alaris GW 800 Volumetric Pump will not operate unless a flow sensor is connected.
Serial Number
Software Version
Configured by
Date
Approved by
Date