Power On/Off
To power up the Pump:
1 . Press the
a
button once and release .
2 . Check:
• A high pitch sounder is activated for 3 seconds during this time the main speaker 'beeps' once upon power up sequence .
• All display segments and all indicators are lit .
• If an error occurs during self-test the Pump will alarm .
3 . After this self-test the Pump will display last rate setting entered or zero, depending on the configuration .
To power down the Pump:
1 . Press and hold
a
. The Pump will display
OFF3
-
OFF2
-
OFF1
.
2 . If the
a
button is released during the countdown the Pump will not power down and it will return to its previous state .
w
If the Pump alarms, segments / indicators do not light up correctly or 2 audible sounds are not heard, then the Pump
must be powered down immediately, and Qualified Service Personnel contacted. If transportation to an engineer is
necessary, it is recommended to use the original protective packaging.
Battery Operation
The Pump will operate from the internal battery when AC Power is disconnected .
The following can be observed to confirm that the Pump is running on battery Power:
• A single beep is heard when AC Power is disconnected from the Pump
• The AC Power Indicator is extinguished
• When infusing:
– The
ml/h
indicator will flash
– Main Display will flash
– Other display indicators, if displaying, will flash then after one minute will extinguish . The display indicators will flash again
after any key is pressed .