BDDF00144 Issue 1
6/40
Alaris™ GW 800 Volumetric Pumps
Controls and Indicators
Controls and Indicators
Controls
Symbol
Description
a
ON/OFF
button - Press once to switch the Pump on . Press and hold down for approximately 3 seconds to switch
the Pump off .
Note:
Logs are maintained for power down events including when the Pump is powered down or
unexpected power loss .
d
RUN/HOLD
button - Press to start the infusion or to put the infusion on hold .
m
CLEAR/MUTE
button - Press to silence alarm or advisories for 1 minute . The audio from the alarm or advisories
will resound after this time . Resets numeric values to zero .
Note:
To re-enable the alarm audio press any other button or key to unmute .
b
PRIME/BOLUS
button - Primes the Infusion set . Administers bolus during the infusion .
l
PRIMARY/SECONDARY
button - Switches the Pump between Primary and Secondary infusion modes . (If
enabled) .
e
ENTER
button - Scrolls between rate, time, VTBI and total volume infused (VI) . Enters values for selected infusion
/ configuration parameters . Confirms the rate during an infusion titration .
hf
CHEVRON
keys - Increases or decreases the infusion rate, TIME limit and VTBI . Press and hold to increase the
selection speed . Used to adjust user selectable options
Indicators: (when illuminated)
Symbol
Description
S
AC POWER
indicator - When illuminated the Pump is connected to an AC power supply .
RATE
The Pump is displaying the infusion rate in millilitres per hour (ml/h) .
VTBI
The Pump is displaying the Volume To Be Infused (VTBI) in millilitres (ml) .
VI
The Pump is displaying the Volume Infused (VI) in millilitres (ml) .
TIME
The Pump is displaying the infusion time in hours : mins .
MICRO
The Pump is operating in the Micro Mode . When not illuminated the Pump is in the Standard Mode .
SEC
The Pump is operating in the Secondary Mode . When not illuminated the Pump is in the Primary Mode .
ml/h
(Millilitres / hour) When
ml
is illuminated the Pump displays the rate, VTBI or VI . When the
h
is illuminated the
Pump displays the rate or infusion time .
ALARM
Indicator - When flashing indicates the Pump is in an alarm condition .
MUTE
Indicator - When flashing indicates the Pump audio alarm is muted .