BDDF00144 Issue 1
22/40
Alaris™ GW 800 Volumetric Pumps
User Selectable Options
User Selectable Options
w
To set the user options the Pump must be on
G
or in set-up mode, and the relevant user selectable options are
enabled. See also 'Configurable Options' section in this DFU.
The Battery Status option is always enabled.
Press
e
button and hold for 2 seconds to enter User Selectable Options .
Note:
Number of
e
button presses are dependent upon the User Select Mode Options which have been enabled . The following
instructions are based upon all options being enabled .
Check the Battery Status
1 . Display shows the status of battery .
•
,
- battery has greater than 30 minutes run time .
•
.
- battery is low and has approximately 30 minutes or less run time .
2 . Press
e
button to return to
G
or set-up mode or next option .
Setting the Occlusion Pressure Level
1 . When
PrES
is displayed .
2 . Select
HI
,
nor
or
Lo
using the
h
f
keys to set the occlusion pressure level to High, Normal or Low .
3 . Press
e
button to return to
G
or set-up mode or next option .
Setting the Alarm Volume
1 . Press
e
button until
tonE
is displayed .
2 . Select an alarm volume between
1
(Low) and
7
(High) using the
h
f
keys .
3 . Press
e
button to return to
G
or set-up mode, or next option .
Setting a VTBI / Time Infusion
1 . Press
e
button until
R
is displayed .
2 . Select
On
or
OFF
using the
h
f
keys to turn the VTBI / time infusion setting on or off .
3 . Press
e
button to return to
G
or set-up mode, or next option .
4 . Make sure that the
TIME
indicator is lit if set to on .
Note:
The time increases / decreases in rate dependant units e .g . 10ml @ 99 .9ml/h is 6mins, therefore 0:06 is displayed .
Setting to Micro Mode
1 . Press
e
button until
T
is displayed .
2 . Select
On
or
OFF
using the
h
f
keys to turn micro mode on or off .
3 . Press
e
button to return to
G
or set-up mode, an audible sounder will confirm the status .
4 . Make sure that the
MICRO
indicator is lit if set to on .