Overload protection: 0.6A/250V fuse; 10A/250V fuse. 
Frequency response: 40

400Hz.    Duty Cycle Display: (0.1%-99.9%). 

XII. Resistance 

1. Press “          “ button, and respectively plug the red and black meter pens   

into the holes of “       ”and “COM”. 

2.Circularly press “          “ button and select resistance measurement.   

Meanwhile, a corresponding symbol will appear on the LCD screen.   
In automatic resistance measurement mode, you could select manual   
measurement by pressing "Range". 

Caution

a) When measuring the capacity, all the powers within the tested circuits must be disconnected and the capacitance shall be 

sufficiently discharged.     

b) When measuring the resistance, any occurrence of voltage may Lead to inaccurate readings, and if the 250V protection 

voltage is exceeded, it may damage the meter or threaten the safety of the user.   

c) When the range of 600

 is in use, first sort-circuit the meter pens and measure the resistance of the lead wires and then   

deduct it in the actual measurement. 

Range 

Accuracy 

Resolution

600

Ω

 

±(0.8%+5d) 

0

.1Ω

 

6k

Ω

 

 

60k

Ω

 

10Ω

 

600k

Ω

 

100Ω

 

6M

Ω

 

±(0.8%+4d) 

1kΩ

 

60M

Ω

 

±(1.2%+10d) 

10kΩ

 

Open circuit voltage: 600mV.   
Overload protection: 250V DC or AC peak values. 

XIII. Diode and On-and-Off Test 

1. Press “        ” button and respectively plug the red and black pens 

 into “     ”and“COM”. 

2. Circularly press “          ” button to select the diode or buzzer measurement. In Diode mode, LDC screen displays "        " 

and voltage symbol,    and the buzzer symbol    "        " is also displayed on the screen.   

3. Connect the red meter pen to the positive of the diode, the black meter pen to the negative of the diode. 

Caution

a) In case the diode is open circuit or the polarities are connected reversely, “OL” will be displayed on the screen. 
b) When checking the diode, all the powers within the tested circuits must be disconnected and the capacitance shall be 

sufficiently discharged. 

c) When the measurement is completed, immediately disconnect the meter and the measured circuit. 

 switch

  10A 

 F

 C

H

z

AC

D

C

/

D

C

/A

C

 F

 C

H

z

AC

D

C

/

D

C

/AC

Summary of Contents for AM-1171

Page 1: ......

Page 2: ...equency temperature diode and make on and off test Meanwhile it is available for unit symbol display data retention the measurement of maximum and minimum value automatic manual range switching automa...

Page 3: ...measuring point and ensure there is no any signal at the input terminal 6 When the electrical measurement is made never let your body get in touch with the ground directly and don t touch uncovered me...

Page 4: ...e minimum value is held After the MAX MIN mode is entered the display device indicates the MAX MIN value No analog bar display and auto power off function in this working environment Press down MAX MI...

Page 5: ...of 61 analog bars 1 4 Measurement mode dual integration A D conversion 1 5 Sampling rate About 3 time s 1 6 Over range Display OL 1 7 Low voltage display about 2 4v the symbol displayed 1 8 Working t...

Page 6: ...Respectively plug in the red and black meter pens into the terminals of and com See the picture below 2 The initial state of the meter is automatic range which shows AUTO symbol Press the Range key a...

Page 7: ...e maximum input current 10A not exceeding 15s Overload protection 0 6A 250V fuse 10A 250V fuse XI AC Current ACA 1 Press the button or button Plug the black meter pen into the hole of COM and the red...

Page 8: ...eter pens and measure the resistance of the lead wires and then deduct it in the actual measurement Range Accuracy Resolution 600 0 8 5d 0 1 6k 1 60k 10 600k 100 6M 0 8 4d 1k 60M 1 2 10d 10k Open circ...

Page 9: ...equency Hz 1 Press button to conduct frequency measurement Respectively plug in the meter pen into V Hz and COM 2 Have the testing end of the meter pen connected in parallel with the signal sources to...

Page 10: ...the terminals of and COM 3 Put the induction end of temperature probe on the surface of the tested object See the picture on the right 4 Read the current testing results from the display Caution a Wh...

Page 11: ...e steps are as follows 1 Loosen the screw on the back cover that secures the battery door and exit the battery door 2 Remove the 3V batteries and replace them with two new ones Although a 3V battery o...

Reviews: