Once the phosphor storage plate has been
inserted into the unit, close the cover and leave
it closed throughout the entire scanning pro-
cess.
The blue status LED lights up.
The image data is automatically transmitted to
the imaging software. The progress of the scan-
ning process is displayed in the preview window
on the monitor.
After it has been scanned, the PSP is erased and
drops into the collection tray.
❯
When the green status LED lights up:
Save the X-ray image.
❯
Remove the PSP and prepare it for recording
another X-ray image.
IDX
a
10.4 Erasing the imaging plate
The image data is erased automatically after
scanning.
The special
ERASE
mode only activates the era-
sure unit of the image plate scanner. No image
data is scanned.
The phosphor storage plate needs to be erased
using the special mode in the following cases:
– The first time the phosphor storage plate is
used or if it is stored for more than one week.
– Due to an error, the image data on the phos-
phor storage plate has not been erased (soft-
ware error message).
❯
Select the special
ERASE
mode in the soft-
ware.
❯
Insert the phosphor storage plate (see "Scan-
ning the PSP").
10.5
Switching the unit off
❯
Press the on/off switch for 3 seconds.
As soon as the unit has shut down it switches
off completely. The LEDs go out.