Plug the enclosed network cable into the net-
work socket of the device.
The cover on the rear must be properly
installed when the device is operated
within the patient environment.
❯
Replace the cover.
8
Commissioning and first
start-up
NOTICE
Short circuit due to the build-up of
condensation
❯
Do not switch on the unit until it has
warmed up to room temperature and
is dry.
The unit supports the following imaging pro-
grams:
– VisionX manufactured for Air Techniques
– DBSWIN manufactured for Air Techniques
– Please inquire about third-party software
Always use the current version of the
imaging program in the commissioning of
the device. Check the version of the
enclosed imaging program versus the ver-
sions available at www.airtech-
niques.com.
8.1
Configuring the network
Network configuration
Various options are available for network configu-
ration:
ü
Automatic configuration via DHCP.
ü
Automatic configuration via Auto-IP for direct
connection of device and computer.
ü
Manual configuration.
❯
Configure the network settings of the device
using the software or, if applicable, the touch
screen.
❯
Check the firewall and release the ports, if
applicable.
Network protocols and ports
Port
Purpose
Service
45123 UDP,
45124 UDP
Device recognition and
configuration
2006 TCP
Device data
514
1)
UDP
Event protocol data
Syslog
1)
The port can vary depending on the con-
figuration.
Installation
24
2160100287L41 1911V006
EN-