Information on EMC according to EN 60601-1-2
13.1
General notes
The information in this leaflet includes excerpts from the relevant European standards for electrical,
medical appliances The information reproduced here should be observed during the installation of in-
dividual appliances and when combining Air Techniques appliances with products of other manufac-
turers If there is any question of doubt, the complete standard must be checked
13.2 Abbreviations
EMC
Electro-magnetic compatibility
HF
High frequency
U
T
Voltage rating of appliance (supply voltage)
V
1
, V
2
Level of consistency for testing according to IEC 61000-4-6
E
1
Level of consistency for testing according to IEC 61000-4-3
P
Rated power of transmitter in watts (W) according to manufacturer's information
d
Recommended safety distance in metres (m)
13.3
Guidelines and manufacturer's information
Electromagnetic transmissions for all appliances and systems
The appliance is designed for operation in one of the electromagnetic environments as outlined be-
low The customer/operator of such an appliance is obliged to ensure that the appliance is operated
in such an environment
Interference
measurements
According
to
Electro-magnetic environment – guidelines
HF transmissions accord-
ing to CISPR 11
Group 1
The appliance employs HF energy exclusively for inter-
nal functions Therefore, any HF transmissions are of
extremely low nature and it is highly improbable that
any other electronic components will receive any inter-
ference
HF transmissions accord-
ing to CISPR 11
Group 2
The appliance must transmit electromagnetic energy in
order to fulfil the functions for which it has been de-
signed Other electronic appliances in the vicinity could
be affected
HF transmissions accord-
ing to CISPR 11
Class [A or
B]
The appliance is designed for use in all types of envi-
ronment including those in residential areas and other
suitable areas which are connected directly to the local
power supply serving residential buildings
Harmonic limits according
to IEC 61000-3-2
[Class A, B,
C, D or Not
Applicable]
Voltage fluctuations/flicker
according to IEC 61000-
3-3
[Fully com-
patible or
not applica-
ble]
Table 1: Electromagnetic transmissions for all appliances and systems