Page 15
Air Techniques, Inc.
9
Instructions for use
NOTICE
Exerting force on the unit can lead to it becoming damaged
• Do not lean against or support yourself using the unit
9.1
Standard settings on switch-on
The tube length can be preset in the Service-Menu "84 Service menu settings"
An extension of the tube influences the image quality and the dose area product displayed
Standard settings on switch-on
– Image plate
– Adult, Pre-molar
– 60 kV, 7 mA
The following table shows the standard values for the image plate exposure time of an adult patient
DC radiator, 7 mA
Tube length 8 in. (20 cm)
DC radiator, 6 mA
Tube length 12 in. (30 cm)
60 kV
µGy
70 kV
µGy
Incisor
008 s
107
013 s
109
Pre-molar
012 s
156
018 s
147
Molar
017 s
215
025 s
207
Bite wing
018 s
227
027 s
223
The following table shows the standard values for the image plate exposure time of a child patient
DC radiator, 7 mA
Tube length 8 in. (20 cm)
DC radiator, 6 mA
Tube length 12 in. (30 cm)
60 kV
µGy
70 kV
µGy
Incisor
005 s
67
008 s
67
Pre-molar
007 s
94
011 s
93
Molar
011 s
146
014 s
144
Bite wing
011 s
146
014 s
144
• Check and adjust X-ray appliances in accordance with the following standard equipment-specific
values
The following table shows the standard values for the sensor exposure times for an adult patient
DC radiator, 7 mA
Tube length 8 in. (20 cm)
DC radiator, 6 mA
Tube length 12 in. (30 cm)
60 kV
µGy
70 kV
µGy
Incisor
007 s
93
011 s
91
Pre-molar
010 s
132
016 s
131
Molar
013 s
162
020 s
164
Bite wing
014 s
174
021 s
172