4 Technical data
Unit electrical data
Nominal voltage
100 to 240 V AC
Max voltage fluctuation
±10%
Frequency
50/60 Hz
Power rating
500 W
Maximum power
750 VA
Classification
FDA 21 CFR Device Classification
Class II
This X-ray system complies with US - FDA:
21 CFR Part 10102
21 CFR Part 102030/31
Degree of protection against ingress of water
Ordinary
Medical products directive (93/42/EEC)
manufacturer: VATECH Co, Ltd
13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggi-do, Korea 445-170
X-ray unit technical data
X-ray Tube Type
Toshiba D-041SB (Stationary Anode Type)
Model
DG-10A05T3
Generator size and weight
55 x 6 x 3 in (140 x 150 x 70 mm)
55 lbs (25 kg)
Source to skin distance
8 in (200 mm) 12 in optional (300 mm)
Generator performance
05 kW
Nominal voltage (Accuracy)
50 to 70 kVp (±10%)
Values below 60 kV are not intended for
human use in USA and Canada
Nominal current (Accuracy)
4 to 7 mA (±20%)
X-ray tube cooling
Automatically controlled
≥
50 °C
Air cooling: optional
Inherent filtration
10 mm Al
Total filtration
Min 20 mm Al
Focal spot size according to IEC 60336
04 mm
X-ray field collimation
∅
24 in/12 x 16 in
2
(
∅
60 mm/30 x 40 mm
2)
Optional 08 x 12 in
2
(20 x 30 mm
2
)
Exposure time (Accuracy)
004 to 2 seconds (±5% or ±20 ms)
Anode material
Tungsten
Anode angle
125 degrees
Duty cycle
1:60 or greater
Product description