Page 13
Air Techniques, Inc.
Setup
Only fully-qualified or from Air Tech-
niques trained personnel may set-up,
install or operate this appliance
6 Prerequisites
The room chosen for set up should fulfil the
following requirements:
Closed, dry room
– Should not be a room made for another pur-
pose (e g boiler room or wet cell)
– No large fields of interference (e g strong
magnetic fields) present, that can interfere
with the function of the unit
– Take environmental conditions into consider-
ation section 4, Technical data"
7 Tube and Power Setup
7.1 Mount tube (optional)
• Place the 12 in (30 cm) tube extension in po-
sition and check whether the grooves match
correctly
1
2
7.2 Mount radiation field
collimation
• Slide the radiation field collimator piece onto
the tube and adjust according to the image
plate or the sensor
1
2
0.8 x 1.2
or
1.2 x 1.6
7.3 Safety for electrical connection
• The appliance may only be connected to a cor-
rectly installed electrical socket outlet
• Do not lay multi-socket units on the floor Speci-
fications of IEC 60601-1-1 apply
• Other systems should not be plugged into the
same multi-socket unit
• All cables, etc must be laid so that no mechani-
cal tension is exerted on them
• Before start-up or first use, check the mains
voltage against the voltage indicated on the
model identification plate (see also "4 Technical
Data")
7.4 Connecting the appliance to the
power supply
Requirements:
9
A correctly installed socket-outlet in the vicinity
of the unit must be available (maximum length
of mains cable 6 ft (18 m)
9
The socket-outlet must be easily accessible
9
Voltage is identical to that given on the model
identification plate
9
Insert plug into electrical socket