Air Techniques, Inc
.
Page 22
Annex
12
Leakage Dose Test Report
12.1
General notes
The X-ray does data is extracted from the X-ray Dose Test Report for the ProVecta HD The X-ray
doses of ProVecta HD in the test report were measured in accordance with the IEC collateral stan-
dards ProVecta HD was designed in accordance with Part 1 Generatl Requirements for Safety, IEC
60601-1-3
12.2
Leakage Dose Test
Test Equipment Information
Instrument
Manufacture
Model
S/N
Dose Meter
Victoreen
660
101114 / 101377
Test Condition
Distance from focal point [m]
1
Applied tube Voltage Peak [kVp]
70
Applied Tube Current [mA]
60
Applied Exposure time [s]
05
Direction [°]
Basic Cone [mGy/hr]
Transverse
Longitudinal
0
00018
00605
45
00140
00395
90
00482
00535
135
00105
00561
180
00035
00947
225
00105
00561
270
00544
00614
315
00702
00509
The tested area on the plane of radiation is defined within a radius at a distance of 1m from the focal
spot