Tips for Operators and Technicians
Repairs above and beyond simple maintenance may only be carried out by a qualified techni-
cian or one of our service technicians
Problem
Probable cause
Solution
E01
Electrical connection between the
control panel and PCB (main
board) is interrupted
• Switch the appliance off
• Check the connection cable
• Inform service technician
E02
Electrical connection between the
x-ray unit and PCB (main board)
is interrupted
• Switch the appliance off and on
• Inform service technician
E03
Current strength exceeded maxi-
mum permitted value during ex-
posure
• Switch the appliance off and on
• Inform service technician
E08
X-ray unit temperature too high
• Switch off the x-ray unit until it
has cooled down sufficiently
• Switch the appliance on again
E09
Current strength outside permit-
ted value during exposure
• Switch the appliance off and on
• Inform service technician
E10
X-ray exposure not activated de-
spite key being pressed
• Quit by pressing key "Menu x-
ray parameter"
• If the fault repeats, inform your
Service Technician
E11
Exposure continues more than
05 secs, even though activation
switch no longer pressed
• Switch the appliance off
• Inform service technician
E12
Displayed kV-value is lower than
the value actually set
• Switch the appliance off and on
• If the fault continues to be
displayed, inform your Service
Technician
E13
Displayed kV-value is higher than
the value actually set
• Switch the appliance off and on
• If the fault continues to be
displayed, inform your Service
Technician
E14
Displayed mA-value is lower than
the value actually set
• Switch the appliance off and on
• If the fault continues to be
displayed, inform your Service
Technician
E15
Displayed mA-value is higher than
the value actually set
• Switch the appliance off and on
• If the fault continues to be
displayed, inform your Service
Technician
Troubleshooting