5 - Rear Panel
Connections overview
11
TGR2050 Series Instruction Manual
⑤
External trigger in
DC coupled External Trigger Input signal, nominal threshold 1.65V
CAUTION
Do not apply external voltages exceeding ±10V to this socket.
⑥
Mod in/out
Analogue MOD IN, AC coupled 10k
Ω
input impedance, 1Vp-p full scale. Analogue
MOD OUT, DC coupled 150
Ω
source impedance, 2Vp-p full scale unloaded. Digital
MOD IN, DC coupled 10k
Ω
input impedance, nominal threshold of +1.65V. Digital
MOD OUT, DC coupled 150
Ω
source impedance, logic levels of <0.8V & >3V
unloaded.
CAUTION
Do not apply external voltages exceeding ±10V to this socket.
⑦
LAN
The LAN interface is designed to meet the LXI standard version 1.5 LXI Device
Specification 2016..
Remote control using the LAN interface is possible using the TCP/IP Socket protocol.
⑧
USB
The USB port accepts a standard USB cable. If the USB driver has been installed from
the website, the Windows plug-and-play function should automatically recognise
that the instrument has been connected. See the support page on the website for
information on installing the driver on a PC at
⑨
GPIB (optional)
IEEE-488 The default GPIB address is 11.