Data entry
To change the arithmetic sign
< +/- >
To select the next position
►
The cursor blinks below the second digit.
P1: 0-10V: 10.000 V
To move back to the previous digit
◄
Each position is programmed like the first.
▲
/
▼
...,
►
...
To complete data input
PROG
To cancel programming
<ESC>
If a data cable or interface connector is plugged in at socket A1 or
DC, key operation on the simulator is - for safety reasons - automati-
cally blocked.
The following should appear in the softkey bar:
Remote Control ! *ON
9. MENUS
The simulator is operated via the main menu and a series of associated sub-
menus in which its operating parameters can be individually set.
9.1 Main menu
In the main menu all parameters for output
ports P0 to P3 can be selected and set directly
by means of keys
PROG
,
▲
/
▼
. see 8.2.
The threshold for the continuity check can be
programmed via port P4. The ports can be
configured in detail via the appropriate sub-
menus.
The 5 Pt100 values can also be selected by
means of.
<SET>
9.2 Submenus
Passing via the main menu to the submenus all ports P0 to P4 can be config-
ured in detail and simulation functions can be activated.
To call up these submenus press
F
►
To return to the main menu press
F
◄
9.2.1 Pt100 output
At socket P0 there are 5 resistors in 4-conduc-
tor circuitry for simulating Pt100 sensors.
(see 6.)
In submenu P0 Pt100 you can by means of
keys
▲
/
▼
(
abbreviation S for steps
) scroll up
and down between the 5 temperature values 0,
50, 100, 200, and 300 °C.
ALMEMO
®
KA7531
9
P0: Output Pt100
Function: steps individually
100.0 °
C
Fª S *ON
SIMULATOR KA 7531 V6.01
P0: Pt100
100.0
°C
P1: 0-10 V
5.000
V
P2: 0-20 mA 12.000 mA
P3: 0-4000Hz 1000. Hz
P4: Continuity yes
36 mV
ESC F ©F SET