Table 2. For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
Infrared thermometer AGU IHE3
is intended for use in the electromagnetic
environment specified below. The customer or the user of
Infrared
thermometer AGU IHE3
should assure that it is used in such environment
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV (contact)
±8 kV (air)
±6 kV
(contact)
±8 kV (air)
Floor should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV air for input/
output lines
Not
applicable
Mains power quality should be
that of a typical commercial or
hospital environment
Surge
IEC 61000-4-5
±1 kV (differential
mode)
±2 kV (common
mode)
Not
applicable
Mains power quality should be
that of a typical commercial or
hospital environment
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip
in U
T
for 0.5 cycle);
40 % U
T
(60 % dip
in U
T
for 5 cycles);
70 % U
T
(30 % dip
in U
T
for 25 cycles);
5 % U
T
(>95 % dip
in U
T
for 5 sec)
Not
applicable
Mains power quality should be
that of a typical commercial or
hospital environment.
If the use of
Infrared
thermometer AGU IHE3
requires continued operation
during power mains interruptions,
it is recommended that
Infra-
red thermometer AGU IHE3
be
powered from an uninterrupt-
ible power supply or a battery
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 А/m
3 А/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment
NOTE:
U
T
– is the AC mains voltage prior to application of the test level.
Summary of Contents for Eaglet IHE3
Page 47: ......
Page 71: ...71 RU AGU IHE3 AGU IHE3 1 1...
Page 72: ...72 C 93 42 EEC 020 2011 WEEE BF 2 12 80 2 15 2...
Page 73: ...73 RU AGU IHE3 1 AGU IHE3 1 2 CR 2032 1 3 1 4 1 3 4...
Page 75: ...75 RU 15 C 35 C 59 F 95 F 15 95 10 6...
Page 76: ...76 30 7...
Page 77: ...77 RU 30 Lo 8...
Page 78: ...78 1 2 3 3 V CR2032 4...
Page 79: ...79 RU 30 10 9...
Page 80: ...80 1 2 3 4 5 6...
Page 81: ...81 RU 7 37 5 C 37 5 C 5 37 5 C 37 5 C 5 8 1 1 2 3 4...
Page 82: ...82 5 6 37 5 C 37 5 C 5 37 5 C 37 5 C 5 7 1 F C F C F 1...
Page 83: ...83 RU 2 3 4 5...
Page 84: ...84 10 1 2 3 10...
Page 85: ...85 RU T 43 C 109 4 F 34 C 93 2 F 15 35 C 59 95 F 11...
Page 91: ...91 RU 2 a AM FM AGU IHE3 AGU IHE3 b 150 80 3 4 AGU IHE3 AGU IHE3 AGU IHE3...
Page 93: ...93 RU agu baby com...
Page 94: ......
Page 95: ......
Page 96: ......
Page 100: ...AGU is the registered trademark by Montex Swiss AG Tramstrasse 16 CH 9442 Berneck Switzerland...