Installation-, Operating-, Maintenance-Manual, Spare Part List
BCR04
AGT-PSG GmbH & Co.KG
Errors and printing mistakes excepted
Page 13
Translation - Manual BCR04 Rev. 6/2017
Components
Condensate-Discharge/Pump (optional)
The sample gas is cooled down to +3°C in the heat exchanger. Water vapor condenses along the dew
point line as a result of the cooling. Condensate is separated from the sample gas and collected in the
condensate separator of heat exchanger. Condensate is continuously discharged by the connected
condensate pump (3,4ml/min @ 5U/min - N4,1x1,6mm).
FEATURES SR25
Self-suctioning peristaltic pump
Reliable continuous condensate removal
Synchronuous motor, gearbox with return stop
Slow speed, hose with long service life
Can operate in any installed position
Electronic Control Unit (optional)
The electronic control unit serves moisture sensors and light barriers of flowmeters and analyses their
signals. The electronic control unit is equipped with a potential-free directional contact, which triggers an
alarm notification or which switches e.g. a sample gas pump or a shut-off valve. The switch-point of the
electronic control unit can be adjusted by a potentiometer. Standard setup: Function type FS 0,5s release
delay, responsivity 75 K
Ω
.
FEATURES EC72.01
Control/alarm for moisture sensor
Control alarm for flowmeter light barrier
Potential-free directional contact
Adjustable switch-point
LED for operation/alarm notification