Installation-, Operating-, Maintenance-Manual, Spare Part List
BCR04
AGT-PSG GmbH & Co.KG
Errors and printing mistakes excepted
Page 7
Translation - Manual BCR04 Rev. 6/2017
Assembly and Installation
Installation Site
The main dimensions are shown at the dimensioned drawing of the sample gas cooler. The weight is
mentioned on the model identification plate.
Failure to install the sample gas cooler in the proper ambient conditions will affect the sample gas coolers
ability to condense refrigerant gas. This can cause higher loads on the compressor, loss of cooler
efficiency and performance, electrical component failure and cooler failure due to the following:
Compressor loss and electrical component failure. Failures of this type will affect warranty considerations.
The layout of the compact sample gas conditioner guarantees a problem-free integration into the analysis
systems.
Minimum installation requirements:
Select a clean dry area, free from dust, and protected from atmospheric disturbances. Avoid
direct solar radiation.
Minimum ambient tempe5°C, maximum ambient tempe45°C.
Specific humidity: 20-70%, non-condensing.
Console mounting mandatory. The mounting rails on the side panels are only for
securing against slipping on the console. Wall mounting with the rails only is not
possible.
With the mounting rails the cooler can be additionally fixed to the wall. The wall has to be perfect
vertical. Ideally you use the optional console-set of the manufacturer or a comparable
construction. The console/brackets must hold 4 times the given weights. Only use the existing
fastening points at the bottom of the housing. The dryer has to be mounted in perfect upright
position.
Ensure unhindered cooling air circulation while installing the sample gas cooler. Do not install the
sample gas cooler directly over, under or beneath possible sources of heat.
CAUTION! Ambient conditions!
Do not install the cooler in areas of extreme dust and dirt or
in corrosive environment.
CAUTION! Min. 80mm free space for ventilation and bleeding!
The required free space at
side- and top-plates for ventilation and bleeding the sample gas conditioner must be maintained
for smooth functioning.
Console mounting mandatory. Ideally you use the optional console-set of the manufacturer or a
comparable construction. For an operation in enclosed housings, such as analytic cabinets, we
recommend a suitable ventilation or air conditioning.
The sample gas cooler is only to lift or move at its bottom panel (2 persons needed). Handle
with care. Heavy blows could cause irreparable damage.
WARNING! Possible damage!
Handle with care. Heavy blows could cause irreparable
damage.
WARNING! Possible damage!
Protect the display during installation of the cooler or during
mounting work. The display is very fragile.